SparkFun Electronics will be closed in observance of memorial day on Monday, May 29th. We will resume normal business hours on Tuesday, May 30th. Any orders placed after 2pm Mountain Time on Friday, May 26th will process and ship out on Tuesday, May 30th.

jacobfb

Member Since: March 15, 2010

Country: United States

  • The sensor soldered on the breakout board needs to be reconditioned for accurate readings. From the SHT15 datasheet:

    IMPORTANT: After soldering the devices should be stored at >75%RH for at least 12h to allow the polymer to re-hydrate. Otherwise the sensor may read an offset that slowly disappears if exposed to ambient conditions.

    I did this by leaving my SHT15 in a ziploc bag with a cigar humidifier rated at rh 84%. You could also us a baggy with a small cup of damp salt. Before reconditioning the SHT15’s readings were -7% what they should be. Now it’s perfect. This is something I wish Sparkfun would do in house.

No public wish lists :(